RTI uses cookies to offer you the best experience online. By clicking “accept” on this website, you opt in and you agree to the use of cookies. If you would like to know more about how RTI uses cookies and how to manage them please view our Privacy Policy here. You can “opt out” or change your mind by visiting: http://optout.aboutads.info/. Click “accept” to agree.
Pulsed helium introduction into a quadrupole ion trap for reduced collisional quenching during infrared multiphoton dissociation of electrosprayed ions
Boue, SM., Stephenson, J., & Yost, RA. (2000). Pulsed helium introduction into a quadrupole ion trap for reduced collisional quenching during infrared multiphoton dissociation of electrosprayed ions. Rapid Communications in Mass Spectrometry, 14(15), 1391-1397.
Previous infrared multiphoton dissociation (IRMPD) experiments utilizing a quadrupole ion trap mass spectrometer yielded limited photodissociation efficiencies. Helium buffer gas continuously infused into the analyzer region at pressures of typically 1 x 10(-3) Torr to improve ion trap performance can collisionally quench photoexcited ions during the IRMPD process. Photodissociation experiments have indicated that uncorrected pressures below 2 x 10(-5) Torr are necessary to avoid collisional deactivation of photoexcited ions. This paper describes IRMPD in the quadrupole ion trap at reduced pressures utilizing a dual-pulsed introduction of helium buffer gas incorporated into the ion trap scan function. The pulsed introduction of helium buffer gas before ion injection allows the efficient trapping of ions injected from an electrospray source and the removal of helium before laser irradiation. A second pulse of helium directly before ion detection improves the intensity of the ion signal. The use of this dual-pulsed inlet of helium for improved IRMPD is demonstrated with the carbohydrate antibiotics neomycin and erythromycin. Copyright 2000 John Wiley & Sons, Ltd