RTI uses cookies to offer you the best experience online. By clicking “accept” on this website, you opt in and you agree to the use of cookies. If you would like to know more about how RTI uses cookies and how to manage them please view our Privacy Policy here. You can “opt out” or change your mind by visiting: http://optout.aboutads.info/. Click “accept” to agree.
Evaluation of a portable Fourier transform infrared gas analyzer for measurements of air toxics in pollution prevention research
Wright, R., Howe, G., & Jayanty, R. (1998). Evaluation of a portable Fourier transform infrared gas analyzer for measurements of air toxics in pollution prevention research. Journal of the Air and Waste Management Association, 48(11), 1077-1084. https://doi.org/10.1080/10473289.1998.10463764
A portable Fourier transform infrared gas analyzer with a photoacoustic detector performed reliably during pollution prevention research at two industrial facilities. It exhibited good agreement (within approximately 6%) with other analytical instruments (dispersive infrared and flame ionization) when analyte concentrations were high and relatively steady. It did not show good agreement when analyte concentrations were low (approximately 10 parts per million [ppm]) or were varying rapidly (less than 1.5 min). The precision for total acetates measurements was estimated to be approximately 40 ppm for measurements in the 0- to 700-ppm region. The precision for styrene measurements was estimated to be approximately 10 ppm for measurements in the 0- to 90-ppm region.